When using ali-Q 2 as a conventional pipet controller, the accuracy and precision (CV) is determined by the graduations on the serological pipet used, which are typically ±1-2%, as well as the user’s ability to consistently eyeball the meniscus and control the liquid dispense.
When using ali-Q 2 as an aliquoting pipet controller, accuracy and precision (CV) specifications are as follows:
ali-Q 2 and ali-Q 2 VS (at speed 7)
Volume Accuracy Precision (CV)
5.0ml +/- 3% +/- 2%
0.5ml +/- 4% +/- 3%
ali-Q 2 Low Speed
Volume Accuracy Precision (CV)
3.0ml +/- 4% +/- 3%
0.3ml +/- 5% +/- 4%
VistaLab Technologies